2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 543
OVER- 3.0.0 ISOLATE INLET WATER PRESSURE (Slow Overflow)
a)
IMPORTANT!
Turn the water OFF!
b) This procedure uses a psi pressure gauge.
ENSURE it
reads 0 psi before installing it.
c) Per the Figure below, install the gauge at Inlet Pressure Regulator #61.
Figure 100 – Inlet Pressure Regulator #61
d)
IMPORTANT! Turn the water ON!
e) Place the machine into RINSE and allow one (1) minute!
f) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
g) ENSURE
Flow Error
= ‘0’ ALWAYS!
h) Is gauge pressure cycling to a Peak of between 18 and 20 psi?
Yes
Between 18 and 20 psi! See procedure number
(page 544).
No
a)
Loosen Regulator’s #61’s lock nut.
b)
Can the center screw be
adjusted until a PEAK between 18 and 20 psi is achieved? TWO (2) possible scenarios:
1)
IF (and ONLY if) between 18 and 20 psi is achieved!
See procedure number
(page 544).
2)
IF between 18 and 20 psi CANNOT be achieved:
TWO (2) possibilities:
1)
Incoming
water pressure more than 105 psi OR;
2)
Rebuild Regulator #61 (Rebuild kit, P/N
190934)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...