2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 514
BL- 3.0.0 AIR SEEN
Per the Figure right, are air bubbles MOVING into the machine
through the clear acid and / or bicarb Inlet Tubing?
Yes Air seen!
A)
If using a jugs ensure they are full AND the
wands reach the bottom of the jugs;
B)
Check the acid
and / or bicarb connector’s O-rings and plug;
C)
Check
the pickup wand(s) tubing.
No air seen! See procedure number
(page 514).
BL- 3.0.1 ISOLATE AIR REMOVAL SYSTEM
a) Turn the water off and allow a “No Water” alarm to occur!
b) The Deaeration gauge is used next.
ENSURE
it reads 0 inHg
before installing it!
c) Allow one (1) minute for the deaeration motor to stop running!
d) Figure right, tee the gauge to the INPUT (clear tubing) of the
Deaeration Pump.
e)
IMPORTANT!
Turn the water on and allow the “No Water”
alarm to go away!
f) Is Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes Deaeration Pressure is OKAY! See procedure number
(page 514).
No Deaeration Pressure in NOT OKAY! ENSURING a “No Water” alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
page 498
SECTION 13 - DEAERATION PROBLEMS
BL- 3.0.4 DEAERATION PRESSURE OKAY
a) Turn the water off and allow a “No Water” alarm to occur!
b) Allow one (1) minute for the deaeration motor to stop running!
c) Remove the gauge and reattach the tubing.
d)
IMPORTANT!
Turn the water on and allow the “No Water alarm to go away!
(page 515).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...