2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 508
UF- 5.0.0 CHECK UF PUMP FOR ‘OPEN CIRCUIT’
a) Turn the machine OFF!
b) Per the Figure right, ENSURE the UF Pump is
plugged PROPERLY into distribution board
position “P22, UF-P”
!
c) Trace the UF Pump’s wiring to distribution board
position “P22, UF-P” looking for insulation
damage.
d) Return to Dialysis Program
(“Select Program” →
‘Dialysis’ → ‘CONFIRM’)
e) Allow two (2) minutes BEFORE continuing.
f) THREE (3) checks. Answer for all three below Check #3:
Check #1:
Does the “UF Pump Alarm” banner appear?
Check #2:
Call debug screen 0. Is the UF Pump symbol pink?
Check #3:
Is the UF Pump’s ‘dot’ ALWAYS blue? (i.e. NEVER white)
Yes (to any of the three):
See procedure number UF- 5.1.0 (page 508).
No (to ALL three):
The problem is no longer present
HOWEVER
:
1)
Check the UF Pump per the
Calibration Procedures AND
2)
Perform the Alarms and Pressure Holding
Tests AND the UF Pump Integrity Test as described in the PM booklet!
NOTE:
If a “UF Pump Alarm” banner reappears intermittently see procedure
(page 508)
UF- 5.1.0 ISOLATE UF PUMP CIRCUIT
“UF Pump Alarm” OR pink UF Pump symbol OR UF pump ‘dot’ is ALWAYS blue:
One at a time swap in the listed components (see COMPONENT LIST below) but in between, to check if the
new component fixes the problem, return to (ABOVE) procedure number
COMPONENT LIST:
1)
UF Pump
1
;
2)
Actuator-Test Board;
3)
Actuator cable
2
;
4)
Distribution board;
5)
Motherboard.
1
With a known good UF Pump, attach all tubing but there is no need to mount it to the hydraulic
compartment unless it checks good.
2
The cable can be checked.
NOTE
that four (4)
UF PUMP
connections will be checked and proceed
to
page 521
SECTION 16 - CHECKING THE ACTUATOR BOARD CABLE
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...