2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 349
CO- 4.9.0 ISOLATE THE ACID / BIC PUMP(S)
a) Per the Figure below, ENSURE no tubing restriction at the Acid and /or Bicarb pump.
Figure 60 – Acid and Bic Pumps
b) At the suspect pump(s), Per the Figure (right), TWO checks:
Check #1:
The metal plate has arrows to identify the Input
(←) and
Output
(→) ports. ENSURE the colored nozzle is at the
Output port AND is towards the TOP of the machine
Check #2:
ENSURE the Input (clear) and Output (solid) tubing is
connected to the appropriate nozzles per check #1.
c) Per the Figures right, disassemble the suspect pump(s). SEVEN
more checks:
Check #3:
ENSURE the seals are oriented correctly.
Check #4:
Check for debris and for broken or bent springs.
Check #5:
ENSURE BOTH springs are ‘weak’ type.
Check #6:
Compare BOTH springs to new and replace one that
appears shorter than a new one.
Check #7:
Check or replace the seals and O-rings.
Check #8:
Check for a torn diaphragm.
f) If a problem was not located above, TWO (2) possible
bad components:
1)
Bad Acid or Bicarb pump OR;
2
)
Bad Actuator-Test Board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...