2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 93
F- 9.0.23 LOW FLOW PUMP PRESSURE / ISOLATE VALVE #43
a) Still in RINSE,
, DOUBLE CLAMP Valve #43’s OUTPUT tubing!
NOTE:
Valve #43’s output tubing extends towards the front of the machine!
Figure 22 - Isolate Valve #43
b) Does gauge pressure NOW CYCLE to more than 35 psi about every three (3) seconds (Yes or No)?
Yes
More than 35 psi! This may indicate Valve #43 is sticking open.
A)
Release the clamps;
B)
Allow Valve #43’s ‘dot’ to turn blue then white again;
C)
While white, if pressure again
does NOT cycle to more than 35 psi then TWO (2) possible bad components:
1)
Bad
Actuator-Test Board
1
OR;
2
) Bad Valve #43.
1
To LOCATE the board refer to
(page 9).
No Less than 35 psi! Leaving the clamps in place, see procedure number
(page 94).
LEFT BLANK INTENTIONALLY