2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 461
TMP- 2.3.4 PDIAL HOLDS / ISOLATE BALANCING CHAMBER VALVES
a) A procedure, in different Section of the Guide, is performed next.
NOTE
this page and procedure number
as you may prompted to return to here.
b) BEFORE continuing to part c, proceed to
page 527
, to perform
SECTION 18A – DIAGNOSTIC VALVE
c) If a leaking Balancing Chamber Valve was not located in part b, was the DiaSafe
®
filter replaced in this
troubleshooting session?
Yes
The DiaSafe
®
filter was already replaced! See procedure number
(page 461).
No The DiaSafe
®
filter was NOT replaced in THIS session! See parts a THROUGH c below:
a) Replace the DiaSafe
®
filter.
b) Place the machine in RINSE for FIVE (5) minutes.
(page 461).
TMP- 2.3.5 DIASAFE FILTER WAS REPLACED PREVIOUSLY
a) Return to Dialysis Program
(“Select Program” → ‘Dialysis → 'CONFIRM’).
b)
IMPORTANT!
From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’!
c) Allow [
Conductivity]
to stabilize to more than 13.0 THEN, ENSURING [Dialysate Flow] has remained
at 500 ml/min, the external flow indicator’s ‘bob’ is rising and falling in the sight tube.
d) Call debug screen 1 to see
FILACT
. It must REMAIN = 0 before continuing to part e!
e) If necessary, reset a TMP alarm up to twice, allowing thirty (30) seconds between each reset, BEFORE
continuing to part f
f)
TMP
is STABLE if (and ONLY if) the TMP window REMAINS white AND TMP does NOT change more
than 20 mmHg in one (1) minute. TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is STABLE:
The Troubleshooting Guide cannot locate an immediate
problem!
2) IF TMP is UNSTABLE:
Assuming ALL procedures were performed correctly AND assuming a
debug screen’s
Flow Error
is NEVER, EVER = 1, Dialysate Pressure Transducer #9* may be bad
intermittent.
*
To LOCATE Transducer #9
refer to
(page 19).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...