2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 430
FIL- 3.7.0 FILACT = 1 CONSTANT OR INTERMITTENT
READ THIS PROCEDURE before performing it:
A)
To avoid damage turn the machine OFF!
B) SEVEN (7) possible bad components (see Component List below). One at a time, swap in each with
known good then in between, continue with parts C through H to test each new component.
Component List:
1)
CBE board;
2)
Sensor Board;
3)
Sensor Board cable;
4)
Functional Board
;
5
) Chamber #69’s cap (Figure right);
6)
Distribution board;
7)
Motherboard
C) When it is time to swap in the Sensor or Functional board, to prevent
“Cond Offset” Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
)).
D) When it is time to swap in Chamber #69’s cap BE SURE to attach the Air
Sensor’s Connector wires to Chamber #69’s probes to the new cap.
E) Turn the machine on and return to Dialysis Program (“Select Program
→‘Dialysis’ → ‘CONFIRM’).
F) From the Home screen, allow [
Conductivity
] to increase to at least 13.0 mS.
G) Allow thirty (30) seconds BEFORE continuing to part H!
H) Call debug screen 1. Watch
FILACT
for two (2) minutes. If EVER = 1 return to part A and swap in the
next component in the list. When
FILACT
remains = 0 the last component swapped in is the problem.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...