2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 436
FIL- 5.5.0 -250 MMHG HOLDS
a) Call debug screen 0.
b) Figure right, if (and ONLY if) Valve #25’s ‘dot’ is BLUE continue
to part c. If NOT blue return to (ABOVE) procedure number
(page 433).
c) Figure right, look at screen 0’s
PDial
data
window. TWO (2)
possible scenarios:
1)
IF (and ONLY if) PDial AGREES with the external meter /- 35:
Proceed to
page 439
, procedure number
2)
IF PDial DOES NOT agree with the external meter /- 35:
See
procedure number
(page 437).
LEFT BLANK INTENTIONALLY