![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 111](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635111.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 107
F- 12.0.0 CALIBRATE/TROUBLESHOOT DIALYSATE PRESSURE
This is NOT a routine calibration! Follow the instructions exactly to avoid error!
a)
Open the shunt door!
b) Enter
Service Mode → Calibrate Sensors →
Dialysate Pressure. The screen says
“1. Connect
a pressure gauge in line…”.
c) Figure right, connect the Four-Way
Assembly (P/N 150034) to the dialyzer
connectors.
d) ENSURE a transducer protector IS NOT in the ‘to syringe’
tubing segment!
e) Place the Four-Way at dialyzer level!
f) If using a
NEO-2
attach to the +Port (top
(red) port). If using a
90XL
attach to the
Pressure Module’s Gauge Port.
g) DO NOT allow tension in the Four-Way’s tubing segments!
h) Clamp the ‘to meter’ tubing segment.
i)
CLOSE the shunt door
and
ENSURE the external flow indicator’s ‘bob’ is moving up and down!
j)
Press ‘CONFIRM’. The screen says “
3. Press [Dialysate Flow on/off] key…”
k) Press the Dialysate Flow on/off key. The Dialysate Flow lamp blinks.
l)
ENSURE the external flow indicator is NOT moving!
m)
Remove the clamp in the ‘to meter’ tubing segment!
n) Using the syringe, adjust pressure until the external meter = between -2 and +2 mmHg.
o) Press ‘CONFIRM’ then see procedure number
(page 107).
F- 12.1.0 PRESSURE TEST
a) The screen says
“6. Pressurize until dialysate pressure reads -250 mmHg…”
.
b) PULL on the syringe plunger. Can you achieve negative (-)250 +/- 5 mmHg on the external meter?
Yes
-250 achieved! Clamp the ‘to syringe’ tubing to hold this pressure then see procedure
number
(page 108).
No -250 CANNOT be achieved! ENSURE the transducer protector, at the meter, is NOT wet OR
consider replacing it! If OKAY, proceed to
page 109,
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...