F- 12.4.0 “OPERATOR ERROR” OCCURED / TROUBLESHOOT DIALYSATE PRESSURE
a) ENSURE the transducer protector, at the external meter, is not WET or consider replacing it!
b) Return the Dialysate lines to the shunt and close the door.
c) Turn the machine OFF!
d) Turn the machine on and return to Dialysis Program (“Select Program”
→ ‘Dialysis’→ ‘CONFIRM’).
e) From the Home screen, select the [Dialysate Flow] window.
f) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
g)
Read before performing!
Return to (ABOVE) procedure number
page 107
) but if (and
ONLY if) “Operator Error” reoccurs, one at a time, swap the listed components (see Component List
below), and in between return to procedure number
page 107
) until “Operator Error” DOES
NOT reoccur indicating the last component swapped in is the problem!
Component List
1)
Dialysate Pressure Transducer #9
*;
;
2)
Sensor Board;
3)
Actuator-Test Board;
4)
Sensor Board cable;
5)
Functional Board;
6)
Distribution board.
*
To LOCATE Transducer #9 see Figure below
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...