2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 169
CLEAN- 2.1.0 ISOLATE INCOMING WATER FLOW
a) Figure right, place the
274 Ω
plug, from the
TWO-RESISTOR SET
, into the float’s
distribution board position, “X5, FLOAT-SW”.
b) Allow forty (40) seconds OR until flow from the
Vent Tubing occurs indicating hydrochambers
A through C are full!
c) MEASURE from the Vent Tubing, for ONE (1) MINUTE. More than eight hundred (800) ml collected?
Yes
More than 800 ml per minute! See procedure number
(page 169).
No Nowhere near 800 ml per minute! Call debug screen 0. Figure
right, is Valve #41’s ‘dot’ BLUE?
Yes ‘Dot’ = Blue!
A)
Reattach the Deaeration Pump’s tubing;
B)
Proceed to
page
No ‘Dot’ = White! a) BE VERY SURE the
274 Ω
resistor plug, from the
TWO-
RESISTOR SET
, is placed PROPERLY in distribution board
position, “FLOAT-SW”! If not, repeat (ABOVE) procedure
number
(page 169)!
b) FIVE (5) possible bad components:
1)
Bad Actuator-Test
Board OR;
2)
Bad Sensor Board cable OR;
3)
Bad Sensor
Board OR;
4)
Bad distribution board OR;
5)
Bad motherboard
CLEAN- 2.1.1 ISOLATE POTENTIAL AIR LOCK
a)
Figure right
, return the float’s connector to distribution
board position, “X5, FLOAT-SW”.
b) Attach the
60 ml syringe
, with the plunger pushed fully
in, to the Deaeration Pump’s clear (INPUT) tubing.
c) PULL on the plunger. If Restrictor #48 is still open, you will
feel very little resistance and can pull the plunger to the
end of the barrel!
d) Remove the syringe and MEASURE flow from the tubing for one (1) minute. If NOT more than 130 ml
repeat parts b through d up to
TEN (10) TIMES
! After no less than ten (10) attempts do measure
more than 130 ml per minute?
Yes More than 130 ml! See procedure number
(page 170).
No Nowhere near 130 ml! TWO (2) possible bad components:
1)
Bad Float Switch* OR;
2)
Bad
Hydrochamber.
* To LOCATE the Float Switch refer to
(page 129).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...