
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 449
TMP- 2.0.3 NO EXTERNAL PUMP LEAKS / ISOLATE FILLING PROGRAM (FILACT)
a) Return the concentrate connectors to acid and bicarb.
b) Return to Dialysis Program (“Select Program”
→ ‘Dialysis’ →
‘CONFIRM)!
c) Call debug screen 1. Is
FILACT
(middle column) = 1?
Yes
FILACT
= 1! Proceed to
page 452
, procedure number
No
FILACT
= 0! See parts a THROUGH d below:
a)
FIGURE BELOW
, inside the distribution, unplug the Air Sensor’s female Connector i.e.
4
th
CONNECTOR,
5
th
position from the left.
NOTE:
If the machine is CBE equipped the CBE
board plugs into the distribution board and Connector plugs into the CBE board which
positions it higher than the other connectors.
b)
Leave the Air Sensor’s Connector unplugged till instructed!
Parts c and d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...