2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 541
I) If
Flow Error
EVER = 1, even just once, a masked Flow
Error is occurring! WITHOUT LOOKING AWAY watch it for
three (3) minutes or until it EVER = 1. TWO (2) possible
scenarios:
1)
IF Flow Error EVER = 1, even just once!
DO NOT troubleshoot overflow instead troubleshoot the
flow error! Refer to the Table of Contents for whatever program the machine is in.
2) IF Flow Error ALWAYS = 0!
Is the machine in:
i)
A Cleaning Program OR
ii )
Dialysis Program:
i)
IF in a Cleaning Program
: Proceed to
page 542
, procedure number
ii)
IF in Dialysis Program:
Perform parts a AND b below:
a) From the Home screen, allow [Conductivity] to increase to more than 12.0 mS.
b) Call debug screen 1. Is
FILACT
ALWAYS = 1 (Yes or No)?
Yes
FILACT
= 1! Proceed to
page 409
No
FILACT
(page 541).
OVER- 1.0.0 ISOLATE CENTRAL (SDS) CONCENTRATE DELIVERY SYSTEM
a)
Turn the machine OFF!
b) When the machine is off OR if [Dialysate Flow] is “OFF” and the machine is attached to a Central
Concentrate Delivery System (SDS), over time, SDS pressure too high may force the concentrate pumps
open causing overflow.
c) Is the red AND / OR blue concentrate connectors currently attached to a SDS system?
Yes
Attached to a central system!
Allowing up to ½ hour
, if (and ONLY if) overflow still occurs see
(page 541). If overflow is NOT occurring maybe the problem
was occurring when attached to a particular station in the clinic where the pressure is too high.
No NOT attached to a central system. If (and ONLY if) overflow is still occurring see procedure
(page 542). If overflow is NOT occurring maybe it was occurring ONLY when
attached to a SDS where the pressure is too high.
OVER- 1.0.1 ATTACHED TO A CENTRAL SYSTEM / ISOLATE HIGH PRESSURE (Slow Overflow)
a) Disconnect the red AND blue concentrate connector(s) from the central supply.
b) Allowing one (1) minute does vent tubing overflow stop?
Yes
Overflow stops! Acid and/or bicarbonate central pressure is too high!
No
Overflow does NOT stop! See procedure number
(page 542).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...