2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 385
CO- 10.0.405 VERIFY SENSOR VERSUS ACTUATOR
Call debug screen 5. TWO (2) checks! Proceed according to the Yes – No choice at check #2.
Check #1 CPRE (middle column) versus ACOND (upper left)):
They
change continually but MUST REMAIN /- 500 of each
other?
Check #2 CPRE versus CPOS (middle column)):
Change continually but
MUST REMAIN /- 500 of each other?
Yes (to both):
CLOSE THE SHUNT DOOR then see procedure number
No (to either OR both):
Proceed to
page 354
, procedure number
CELLS but if (and ONLY if) you return here see procedure number
(page 386).