2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 161
CLEAN- 1.2.33 LOADING PRESSURE REMAINS LOW
a)
Place the machine into RINSE (NOT HEAT DISINFECT)!
b) Call debug screen 0 (Figure below). If debug does not appear press ‘Escape’ then call screen 0.
c) Locate
DEAP
. TWO (2) possible scenarios:
1)
IF (and ONLY if) DEAP = 75 OR less:
2) IF DEAP = 76 OR more:
See parts a THROUGH i below:
a)
Enter Service Mode → Calibrate Hydraulics → Deaeration Pressure.
DO NOT follow the
screen instructions! See part b instead!
b) Press ‘CONFIRM’. The screen’s
[Pump Rate]
window turns yellow.
c) Select the
[Pump Rate]
window to turn it bright yellow!
d) Set the
[Pump Rate]
window to “210”!
e)
Press ‘CONFIRM’ twice to save the calibration!
f) Turn the machine off then back on.
g)
Place the machine in RINSE (NOT HEAT DISINFECT)
.
h) Call debug screen 0. If debug does not appear press ‘Escape’ then call debug screen 0.
i)
Is
DEAP
now = 45?
Yes
DEAP
(page 162).
No
DEAP
NOT = 45! CAREFULLY repeat the Deaeration Pressure Calibration from part
a. After repeating, if (and ONLY if)
DEAP
STILL does not = 45, THREE (3) possible
bad components:
1)
Actuator-Test Board OR;
2)
Functional Board EEPROM (IC2)
OR;
3)
Functional Board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...