2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 572
WD- 1.0.1 ISOLATE CONTROL SIGNAL
a) Per the Figure below, locate IC2 at the top edge of the Power Logic Board.
Figure 108 – Power Logic Board / R10
b) Under IC2 are three resistors, top to bottom R8, R9 and
R10
.
c) Measure at the RIGHT side of
R10
. TWO (2) possible scenarios:
1)
IF (and ONLY if) 3.6 volts DC or more:
The Functional Board is bad (to locate the board refer to
2) IF less than 3.6 volts DC:
(page 572).
WD- 1.0.3 ISOLATE THE POWER LOGIC BOARD
a) Turn the machine OFF and swap in a known good Power Logic Board (to LOCATE the board refer to
b) Turn the machine on. TWO (2) possible scenarios:
1)
IF “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” does NOT reoccur:
Turn the machine off
and on several times, allowing 40 seconds in between, as the problem may be intermittent! If it
does not reoccur the new Power Logic Board fixed the problem.
2)
IF “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” REOCCURS:
: Perform parts a THROUGH d
below:
a) Turn the machine OFF! The previous Power Logic Board is probably okay!
Parts b through d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...