2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 212
VEC- 2.0.0 ‘BOB’ MOVING UP AND DOWN
A) Turn the machine OFF!
B) This procedure checks Valve #39 in RINSE. If it opens, as it should, deaeration pressure will be
between 0 and -13 inHg. See parts a THROUGH e below:
a) The deaeration gauge is here. ENSURE it reads 0 inHg before installing it!!
b) Fluids and tubing may be HOT but won’t scald you!
c) Figure below, tee the gauge into the Inlet (clear) tubing of the Deaeration Pump.
d)
Place the machine in RINSE!
e) Reading the gauge, TWO (2) possible scenarios:
1)
IF (and ONLY if) between 0 and -13 inHg:
Valve #39 is open!
See procedure number
(page 213)!
2)
IF between -13 and -25 inHg:
ENSURING the machine was in RINSE prior to checking
pressure, Valve #39 IS NOT opening! FOUR (4) possible bad components:
1)
Bad Actuator-
Test Board OR;
2)
Bad Valve #39 (see Figure above) OR;
3)
Bad ACTUATOR cable OR;
4)
Bad distribution board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...