BL- 1.0.5 BLOOD LEAK ALARM?
Is the Blood Leak alarm still present (Yes or No)?
Yes
Press and release the ‘RESET’ key up to three (3) times in twenty (20) second intervals. TWO (2)
possible scenarios:
1)
IF the Blood Leak alarm clears:
2) IF the Blood Leak alarm does NOT clear:
(
page 513
No The alarm clears! See procedure number
(page 512).
BL- 1.0.6 BLOOD LEAK ALARM NOT PRESENT (CHECK FOR AIR)
a) RESET ALL alarms and turn the blood pump on to 100 ml/min or more.
NOTE:
If the blood pump will not
run perform the alarms tests.
b) From the Home screen, set
[UF GOAL]
to 1000 ml; set
[UF Time
] to 1:00 hr.
c) Press ‘CONFIRM’.
d) To turn UF on, press the front panel’s UF on/off key. The green on/off lamp
turns on solid.
e) Allow
TMP
to increase to 160 mmHg THEN IMMEDIATELY turn UF off (UF lamp off).
TMP
should
increase to no more than 300 mmHg
1
.
1
If a “Filling Program” banner appears allow it to clear. If “Fillling Program” occurs more than twice an air
leak may be indicated. In this event (ONLY) proceed to
page 514
, procedure number
f)
Per the Figure right,
using a
flashlight
, watch for air bubbles
through the Blood Leak Sensor’s TO
AND FROM tubing for one (1) minute
each.
g) TWO (2) possible scenarios:
1)
IF (and ONLY if) air is seen!
See
(
page 514
).
2)
IF no air is seen:
Proceed to
page 520
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...