2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 542
OVER- 2.0.0 ISOLATE POTENTIAL FLOW RECIRC ERROR
Is the machine currently in Heat Disinfect?
Yes
In Heat Disinfect! See procedure number
(page 542):
No NOT in Heat Disinfect! See procedure number
(page 542).
OVER- 2.0.1 IN HEAT DISINFECT
a) Press ‘Escape’ to return to the Main Heat Disinfect screen.
b) Allow Remaining
Prerinse Time
to = 0:00.
c) Allow up to five (5) more minutes. If a
“
Flow Recirc Error 1” banner appears DO NOT troubleshoot
overflow. If (and ONLY if) a
“
Flow Recirc Error 1” does NOT occur see procedure number
(page 542).
OVER- 2.0.2 ISOLATE HEAT EXCHANGER #77 (Slow Overflow)
a)
IMPORTANT! Turn the machine OFF but leave the water ON!
b) Per the Figure below, remove the front CLEAR tubing (to drain) from the Heat Exchanger.
Figure 99 – Hydraulics Front View / Heat Exchanger Check
c) Watch the vacant Heat Exchanger nozzle for up to THREE (3) FULL minutes. or until you see dripping,
possibly VERY slowly (Yes or No)?
Yes
Dripping seen!
IMPORTANT!
YOU MUST
perform the following TWO procedures:
Procedure #1:
Turn the water OFF and replace the Heat Exchanger;
Procedure #2:
To
prevent damage to the new Heat Exchanger, see procedure number
(page 551)
IMMEDIATELY.
No dripping! Reattach the tubing then see procedure number
(page 543).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...