2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 255
T- 1.5.53 [TEMPERATURE] IS STABLE
WITHOUT LOOKING AWAY (to avoid damage), watch the external meter for up to twenty-five (25) minutes
or until if it exceeds 95° C! TWO (2) possible scenarios:
1)
IF (and ONLY if) the meter exceeds 95° C:
See procedure number
(page 256).
2)
IF the meter does NOT exceed 95° C:
Perform parts A THROUGH I below as prompted:
a) Remove the 80° C resistor plug from the distribution board.
b) Turn the machine off and replace
NTC #3
*
with a known good, and plug it into the
2
nd
distribution
board position from the left, “MON-NTC”.
*
To LOCATE NTC #3 refer to
(page 218)
c)
Read before performing!
Return to (ABOVE) procedure number
(page 247) to see if the
new NTC #3 fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.53 continue
to part D.
d) Turn the machine off and swap in a known good Sensor Board* then continue to part E. *To
LOCATE the board refer to
(page 9).
e) Perform TWO (2) calibrations:
Cal 1)
Temp Control;
Cal 2)
Cond Cells. After the calibrations
continue to part F.
f)
Read before performing!
Return to (ABOVE) procedure number
(page 247) to see if the
new Sensor board fixed the problem. If (and ONLY if) you return to procedure number
T- 1.5.53 continue to G.
g)
Read before performing!
Turn the machine off and swap in a known good Functional Board THEN
see part H.
h) Perform FOUR (4) calibrations:
Cal 1)
Voltage Detection;
Cal 2)
Temp Sensor;
Cal 3)
Temp
Control;
Cal 4)
Cond Cells. After the calibrations see part I.
i)
Return to (ABOVE) procedure number
(page 247) to see if the new board fixed the
problem.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...