2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 39
F- 2.0.15 [TEMPERATURE] NOT 35.0 - 39.0 °C AND / OR [CONDUCTIVITY] NOT 13.0 - 14.5 mS
a) Call the Home screen.
b) Select the [
Temperature
] window to ENSURE “Temp Setting” is 37.0° C.
c) Press ‘CONFIRM’.
d) With the shunt door REMAINING open, call debug screen 0.
e) WITHOUT LOOKING AWAY, watch
Flow Error
for one
(1) minute. If it EVER = 1, even just once indicates a
Flow Error. Is it EVER = 1?
Yes
Flow Error
EVER
=
1! Proceed to
page 47
, procedure number
No
Flow Error
ALWAYS = 0! See parts a AND b below:
a) Call debug screen 6.
b) WITHOUT LOOKING AWAY, watch
BC Switch
(middle column)
for five
(5) full minutes OR until if it EVER = 897 or more, even just once,
indicating a Flow problem. TWO (2) possible scenarios:
1)
IF (and ONLY if) BC Switch is EVER = 897 or more even if only once:
Proceed
to
page 47
, procedure number
2) IF BC Switch is NEVER = 897 or more:
(page 39).
F- 2.0.16 NO FLOW ERRORS
Call the Home screen. THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) the Temperature window is red:
Proceed to
page 214
SECTION 4 – TEMPERATURE PROBLEMS
2)
IF (and ONLY if) Conductivity is NOT between 13.0 and 14.5 mS:
Proceed to
page 317
SECTION 5 – CONDUCTIVITY PROBLEMS
3)
IF the Temperature window is pale yellow/white AND Conductivity is between 13.0 and 14.5
mS:
page 36
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...