2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 105
Yes
PDial
= 0 +/- 50!
A)
Return the dialyzer connectors to the shunt and close the door!
B)
See procedure number F- 11.0.5 (page 105).
No
PDial
is NOT = 0 +/- 50!
A)
IMPORTANT!
Turn Dialysate flow ON (Flow on/off lamp NOT
blinking);
B)
Proceed to
page 107,
F- 11.0.5 TMP ALARM?
a) Turn Dialysate Flow ON (Dialysate Flow on/off lamp STOPS blinking).
b) Call the Home screen.
c) Set [Dialysate Flow] to
500 ml/min
and press ‘CONFIRM’!
d) Is the TMP window RED (TMP alarm)?
Yes TMP window IS RED!
RESET SEQUENCE
:
A)
Press and release the ‘Reset’ key then
immediately press and hold it for three (3) seconds.
B)
Allow thirty (30) seconds.
C)
If a TMP
alarm reoccurs perform the RESET SEQUENCE up to twice more BEFORE continuing to
procedure number
(page 105).
No TMP window is white! See procedure number
(page 105).
F- 11.0.6 ISOLATE FOR A LEAKING BALANCING CHAMBER VALVE
TMP is STABLE if (and ONLY if) the TMP window REMAINS white AND TMP does NOT change more than
60 mmHg in three (3) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is STABLE:
(page 106).
2)
IF TMP is UNSTABLE:
A procedure in a different Section is performed next!
NOTE
this page and
procedure number (F- 11.0.6) as you may be because you may be prompted to return here. See part A
below:
A) Before continuing to part B, proceed to
page 527
B) If a balancing chamber valve leak was not located in part A turn the machine off then see
(page 106).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...