2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 106
F- 11.0.7 CHECK FOR SECONDARY SIDE RESTRICTION
a) Obtain a 1000 ml graduated cylinder.
b) Plug the red concentrate connector into its rinse port.
c) Turn the machine on but
DO NOT
press any keys!
“Select Program” remains up till instructed
otherwise!
d) Remove the red dialyzer connector from the shunt door and place it into a 1000 ml graduated cylinder.
e) Plug valve #37 into valve #30’s distribution board position “V30”. This opens valve #37.
Valve #30
remains unplugged!
f) With “Select Program” up, measure flow, from the dialyzer connector, for one (1) minute. If no flow,
double check valve connector placement
. TWO (2) possible scenarios based on measured flow:
1)
IF (and ONLY if) less than 800 ml collected:
A)
Return the dialyzer connector to the shunt but
DO NOT
unplug valve #37;
B)
page 71
).
2)
IF 800 ml or more collected:
See parts a THROUGH d below:
a) Turn the machine OFF!
b) Return valves #37 and #30 to their distribution board positions.
c) Turn the machine on and return to Dialysis Program (
“Select Program” → ‘Dialysis’ →
‘CONFIRM’)!
d) Call debug screen 0. Watch
Flow Error
for six (6) minutes OR until it EVER = 1! TWO (2)
possible scenarios:
1)
IF Flow Error EVER = 1:
TWO (2) possible bad components:
1)
DiaSafe
®
filter OR;
2)
transducer #10*. *To LOCATE the CFS transducer refer to
(page 92).
2)
IF Flow Error ALWAYS = 0.
The flow error is not occuring at this time!
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...