T- 7.2.0 TEMPERATURE CONTROL CHECKS
a) Return NTC #3’s connector to the
2
nd
distribution board position from the left, “MON-NTC”.
b) ENSURE NTC #2’s connector is in the
1st
distribution board position from the left, “CON-NTC”.
c)
Turn the Heater Switch ON!
d) Install the hydraulics into the cabinet however, there is no need to screw the rear panels in!
e) Enter
Service Mode → Options → Hardware Options. Does the “Yes” box at
‘
Temp Comp’
have an ‘X’ in it?
Yes
Place the ‘X’ in the “No” box and press ‘CONFIRM’. The ‘X’ turns
blue. Post a note that ‘
Temp Comp
’ is off then see procedure
No
Temp Comp
(page 305).