2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 558
LEAKING- 2.0.0 SECONDARY SIDE LEAK / DIALYSATE SAMPLING OPTION?
a) From the Service Mode menu, select the ‘Options’ button.
b) Select the ‘Hardware Options’ button.
c) Does the “Yes” box at
Dialysate Sampling
(Figure right) have a blue ‘X’
in it?
Yes
Dialysate Sampling
(page 558).
No
Dialysate Sampling
= “No”! This is NORMAL! In what Program,
1)
Dialysis OR
2)
A Cleaning
Program (Rinse, Heat Disinfect, etc.) was the leak ORIGINALLY seen?
1)
IF (and ONLY if) in Dialysis Program:
Turn the machine OFF then proceed to
page
, procedure number
2) IF in a Cleaning Program:
Proceed to
page 559
, procedure number
LEAKING- 2.1.0 DIALYSATE SAMPLING = “YES”
a) Place the ‘X’ in the “No” box and press ‘CONFIRM’.
b) ENSURE the “No” box at
Dialysate Sampling
now has a blue ‘X’ in it!
c) Turn the machine OFF then back on!
d) Return to the Program (Cleaning OR Dialysis) where the leak was
ORIGINALLY
seen! In a Cleaning Program squirting into the bucket
occurs if (and ONLY if) UF Check Valve #63 is okay!
e) Figure right, if (and ONLY if) no leaks from the Fluid Sample port = UF
Check Valve #64 is okay!
f) Watch for up to four (4) FULL minutes OR until an active leak reoccurs. TWO (2) possible scenarios:
1)
IF (and ONLY if) an active leak reoccurs:
Turn the machine OFF then see procedure number
page
.
2) IF an active leaks does NOT reoccur AND THREE (3) possible scenarios i) or ii) or iii below:
i) IF in a Cleaning Program AND a Flow Error occurs:
Proceed to
page 158
, procedure
ii) IF in Dialysis Program AND a Flow Error occurs:
Proceed to
page 20
iii) IF a Flow Error does NOT occur:
The leak was probably due to the incorrect Dialysate
Sampling setting! This option should always be left at “No”!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...