2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 407
CR- 1.0.9 ISOLATING NEGATIVE (-) TWELVE VOLTS
a)
IMPORTANT! To avoid damage turn the machine OFF!
b) Figure below, at the rear of the card cage, ENSURE the Blood Pressure Module cable, is NOT reverse
connected with another module!
If it is, correct this situation then re-measure -12 volts as this
may be the problem!
c)
IMPORTANT!
Unplug the
Blood Pressure
module.
d)
IMPORTANT! Turn the machine on (fan running)!
e)
ENSURE the meter’s black lead REMAINS connected to ground!
f) Repeat the measurement at TEST Connector’s
-12V
pin (pin 4, four pins from left). Between -11.0 and
-13.0 volts DC now?
Yes
Between -11.0 and -13.0 DC! The Blood Pressure module OR its cable may be bad.
No NOT between -11.0 and -13.0 DC! See procedure number
(page 407).
CR- 1.0.10 NEGATIVE 12 VOLTS NOT WITHIN RANGE
a)
IMPORTANT!
To avoid damage
turn the machine off THEN allow two (2) full minutes!
b)
IMPORTANT! Set your volt meter to measure resistance
(Ω)!
c)
ENSURE the meter’s black lead REMAINS connected to ground!
Parts d and e next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...