2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 396
NOTE: Table B readings are affected by changing the [Dialysate Flow] rate, or turning it off, Filling
Programs, Pressure Tests and Temperature or Conductivity alarms (i.e. bypass). ENSURE the
machine is stable before responding to variations!
Table B – Debug Screen 5 OLC Values
Parameter
Description
Value
Qd
Dialysate Flow Rate
Should be /- 10 ml/min of the selected Dialysate
Flow rate
QdS
Dialysate Flow Stability
0 =
Dialysate Flow is unstable
1
= Dialysate Flow is stable.
NOTE:
QdS may = 0 if a Temp or Cond alarm occurs
OR if Dialysate Flow rate was changed OR if a Filling
Program occurs OR if a TMP alarm is reset
Current TI
Pre Dialyzer Temp sensor
NTC #3 stability
Ranges from 0 (unstable) to 100 (very stable)
QbS
Blood Pump Stability
Counter
Normally 0 or 1. If QbS reaches 20, the OLC test will
not start OR will cancel
Avg TI
Pre Dialyzer Temp stability
over the last 12 minutes
When Dialysate Flow = 500 ml/min, should be more
than 70%. May be lower at other flow rates.
Pre Stbl Ct
Pre Dialyzer counter
Defaults to 90. Decreases to 0 if Pre Dialyzer Cond is
stable
Pre % Stbl
Percentage of time Pre
Dialyzer Cond was stable
over the last 12 minutes
If OLC test is NOT running should be more than 90%.
Instability (i.e. less than 90%) is NORMAL within 12
minutes of an OLC test.
Post Stbl Ct
Post Dialyzer Cond counter
Defaults to 70. Decreases to 0 if Post Dialyzer Cond is
stable.
Post % Stbl
Percentage of time Post
Dialyzer Cond was stable
over the last 12 minutes
If OLC test is NOT running should be more than 90%.
Instability (less than 90%) is NORMAL within 12
minutes of an OLC test.
OLC status
Current stage of OLC test
0
= An OLC test is NOT running
1
= At or going to approx.15.5 mS
2
= At or going to the approx. 13.5 mS
OLC Enable
OLC Enabled
0
= OLC is turned off (OLC will NOT run).
1
= If OLC is turned on.
CPRE Stable
Pre Dialyzer Conductivity
when stable during OLC
test
Example:
15318 = 15.318
mS
Not valid UNLESS an OLC test was performed.
CPOS Stble
Post Dialyzer Conductivity
when stable during OLC
test
Example:
15318 =
15.318 mS.
Not valid UNLESS an OLC was performed.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...