2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 584
SECTION 24 – “DIASAFE
TEST FAILED”
A) If a DiaSafe
®
test is running allow it to finish BEFORE CONTINUING to part B. It is NORMAL for
Flow
Error
= 1 WHILE a DiaSafe
®
test is running!
B) Using a flashlight, ENSURE no air bubbles flowing into the machine through the Acid and Bicarbonate
inlet tubing.
If air is seen there is a problem with a concentrate connection.
C) From the Home screen, select the [Dialysate Flow] window.
D) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
E) ENSURE a “No Water” alarm NEVER occurs!
F) Call debug screen 0. WITHOUT LOOKING AWAY, watch
Flow Error
for two (2) minutes. If EVER = 1 indicates a Flow
problem. TWO (2) possible scenarios below:
1)
IF (and ONLY if) Flow Error REMAINS = 0!
Continue to part G.
2) IF Flow Error EVER = 1!
Proceed to
page 20
SECTION 1 - FLOW ERRORS IN DIALYSIS
G) From the Home screen, [
Temperature
] MUST be between 35.5 and 38.5° C; [
Conductivity
] between
13.0 and 14.3 mS! BOTH MUST be stable i.e. NOT changing more than 0.2 per minute.
DIASAFE- 2.0.0 FLOW ERROR ALWAYS = 0 / ISOLATE PRESSURE TESTS
a) Using a flashlight, ENSURE no air bubbles through the external flow indicator’s sight tube*!
*
Air is normal after a Diasafe
®
test but should clear within four (4) minutes! Prolonged air cause Filling
Programs (i.e.
FILACT
= 1) and TMP will NOT stabilize!
b)
Reset ALL alarms!
c) At the bottom of the screen, select the
[Test & Options]
tab.
d) If the
[Pressure Test]
key is
gray
select
[Both Tests]
. If blue select [
Pressure Test
]!
e) Press ‘CONFIRM’ and allow the tests to finish!
f) Do BOTH Negative AND Positive Pressure Tests pass?
Yes
BOTH Pressure Tests pass! See procedure number
(page 585).
No A pressure test fails*! Proceed to
page
SECTION 10 – PRESSURE TESTS FAILING
* DO NOT troubleshoot “Diasafe Test Failed” if a Negative OR Positive
pressure test fails!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...