2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 183
CLEAN- 4.6.33 ISOLATE VALVE #43 POSSIBLY STICKING CLOSED
PDIA
is NEVER 6.0 or AND NEVER 0.0 longer than seven (7) seconds!
a) Call debug screen 0 (Figure right).
b) Allow Valve #43’s ‘dot’ to turn blue then white
again.
c) As seen earlier, ENSURE the external Flow
Indicator’s ‘bob’ is rising at least ¼ way up in the
sight tube!
d) Per the Figure below, unplug the Flow Pump from Distribution Board position “P21-Flow-P”.
e)
ENSURE the FLOW MOTOR shaft stops rotating!
f) Every forty five (45) seconds Valve #43’s ‘dot’ turns blue for sixteen (16) seconds. WHILE IT IS BLUE,
does the external flow indicator’s ‘bob’ rise at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! Valve #43 is okay!
A)
Return the Flow Pump’s connector to distribution
board position” P21- Flow-P”;
B)
ENSURE the Flow Motor is running;
C)
If
“Flow Recirc
Error 1”
continues to occur you may have made an error somewhere!
No ‘Bob’ IS NOT moving! ENSURING Loading Pressure (gauge in the rinse port) remains good,
TWO (2) possible bad components:
1)
Bad Actuator-Test Board OR
2)
Bad Valve #43
(sticking closed).
NOTE:
When troubleshooting is complete be sure to return the Flow Pump’s connector to
distribution board; position” P21- Flow-P”;
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...