2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 146
NW- 1.0.10 “NO WATER” ALARM REOCCURS
a) Leaving the machine on, turn the WATER OFF!
b) A psi pressure gauge is required.
ENSURE
it reads 0 psi before installing it!
c) Figure below, tee the gauge into Inlet Pressure Regulator #61.
d) Tie wrap both sides of the gauge tubing to prevent leaks and false readings!
e)
TURN THE WATER ON!
(page 146).
NW- 1.0.12 GAUGE IN AND WATER ON / ISOLATE INLET PRESSURE
Call debug screen 0, to watch Valve #41’s ‘dot’ for forty-five (45) seconds. TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve #41’s ‘dot’ is NEVER white (ALWAYS blue):
If ABSOLUTELY SURE the
water is ON and the ‘dot’ still is NEVER white proceed to
page 148,
2)
IF Valve #41’s ‘dot’ is remaining white OR cycles between white and blue!
TWO (2) possible
scenarios below:
Scenario #1:
IF (and ONLY if) REMAINING white:
If (and ONLY if) a “No Water” alarm NEVER
reoccurs whatever was causing the problem is no longer present! If a “No Water”
alarm does reoccur see Scenario #2 below.
Scenario #2: IF cycling between white and blue:
Gauge pressure should cycle between a PEAK of
more than 18 psi and to NEVER LESS than 8 psi. WITHOUT LOOKING AWAY, watch
the gauge AND the screen two (2) FULL minutes preferably until a “No Water” alarm
reoccurs. TWO (2) possible scenarios 1) or 2) next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...