2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 364
CO- 7.3.0 ISOLATE POSITIVE (+)12 VOLT DC SUPPLY
a) CAREFULLY return Cond Cell #7 to distribution board position “X7, COND”.
b)
Turn the machine OFF
.
c) To avoid pulling cables loose, GENTLY open the card cage.
(page 364).
CO- 7.3.1 I12 VOLT DC SUPPLY
a) Return to concentrate connectors to their rinse ports.
b) Place the machine into RINSE!
c) Spread the card cage side panels open then gently drop the front panel down to access the
motherboard’s nine (9) pin TEST Connector (Figure below).
d)
IMPORTANT! Set your CALIBRATED volt meter to DC volts (V
DC
).
e)
Connect the meter’s black lead to chassis ground (see Figure 2 (page 4)).
f) Per the Figure below, measure at the TEST Connector’s
+12V
pin (pin 5, five pins from the left). TWO
(2) possible scenarios:
1)
IF (and ONLY if) between 11.7 and 13.3 volts DC:
See procedure number CO- 7.3.4 (page
2) IF NOT between 11.7 and 12.3 volts DC:
Replace the Power Logic Board
*
.
*
To LOCATE the board refer to Figure 4A (page 9).