2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 419
FIL- 2.0.4 NO AIR SEEN / ISOLATE FILLING PROGRAM
Call debug screen 1. WITHOUT LOOKING AWAY watch
FILACT
(middle column) for up to two (2) minutes. If EVER = 1, even just once,
indicates a Filling Program! TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT EVER = 1:
Proceed to page
2) IF FILACT ALWAYS = 0:
See parts a THROUGH d below
:
a)
Open the shunt door and LEAVE IT OPEN till instructed otherwise!
b)
Place a ‘dummy venous chamber’ in the Level Detector module.
c) Press and release ‘Reset’ key then immediately press and hold it for three (3) seconds. Allow thirty
(30) seconds. If any alarm reoccurs attempt reset up to twice more BEFORE continuing to part d.
d) Can you RESET ALL alarms?
Yes
All alarms reset! See procedure number
(page 419).
No Address the alarm that cannot be reset!
FIL- 2.0.5 ALARMS RESET
Call debug screen 0. Per the Figure below, WITHOUT LOOKING AWAY, watch the UF Pump’s ‘dot’ for
forty five (45) seconds. Does it EVER BLINK between white and blue?
Yes
The UF ‘dot’ blinks! The Actuator-Test Board may be bad!
No
The UF ‘dot’ does NOT blink! See procedure number
(page 420).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...