MOTORS- 4.5.0 ! EMPTY = 1 / ISOLATE MOTOR BRUSH CONNECTION
a) UNPLUG the NOTED motor from the distribution board!
b) Using a flashlight, check
inside
the motor’s distribution board position for ‘white’ corrosion or damaged
male
pins.
Damage indicates the distribution board may need to be replaced!
c) Figure right, remove the plastic cap from the female
distribution connector.
•
If the
NOTED
motor is the
Deaeration Motor
it
MUST be connected between
pins 1 and 3 i.e. top
and middle
•
If the
NOTED
motor is the
Flow Motor
,
it MUST be
connected between
pins 1 and 5 i.e. top and bottom.
d) Set your CALIBRATED voltmeter to
RESISTANCE
(Ώ)
.
e) Touch the meter’s leads together. The meter MUST read less than 0.4
Ω
! Subtract this from the
subsequent measurement.
f) Measure,
INSIDE
the female connector, BETWEEN the soldered terminals. Less than ten (10)
Ω
?
Yes
Less than 10 Ω! See procedure number
(page 136).
No
More than 10 Ω! Bad brushes or motor.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...