![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 235](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635235.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 231
1)
IF (and ONLY if) the [Conductivity] window turns white:
See procedure number
(page 231)
2)
IF AFTER three (3) minutes the [Conductivity] window remains RED:
Proceed to
page
, procedure number
T- 1.0.19 VERIFY ‘OUT OF BYPASS’
Call debug screen 0. Figure right, is Valve #24’s ‘dot’ blue OR white?
1)
IF (and ONLY if) BLUE:
2) IF WHITE:
See parts a THROUGH d below:
a) Call debug screen 2.
b) To ENSURE the shunt door is CLOSED,
CVRCLS
(2
nd
column
from left) = 1!
c) Call the Home screen to ENSURE the [
Temperature
] AND [
Conductivity
] windows are pale
yellow/white.
d) Return to debug screen 0. Allow up to three (3) minutes OR until if Valve #24’s ‘dot’ turns blue. TWO
(2) possible scenarios:
1)
IF (and ONLY if) the ‘dot’ turns blue:
See procedure number T- 1.0.191 (page 231).
2)
IF the ‘dot’ turns REMAINS white:
Proceed to
page 302
, procedure number
T- 1.0.191 VALVE #24’S ‘DOT’ BLUE / VERIFY ‘OUT OF BYPASS’ CIRCUIT
a) Call the Home screen.
b) Set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
c) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes ‘Bob’ moving! See procedure number
(page 232).
No ‘Bob’ is not moving! See parts a AND b below:
a) From the Home screen, ENSURE the [
Temperature]
AND [
Conductivity]
windows are
pale yellow/white.
b) Call debug screen 0. If Valve #24’s ‘dot’ is still blue AND the ‘bob’ still is not moving there
may be masked Flow Error.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...