![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 636](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635636.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 632
P- 1.0.6 X2 / PIN 3 MORE THAN 6.0 VOLTS / ISOLATE POWER LOGIC BOARD
Continuing to measure at X2,
pin 3
press and release the Power button. TWO (2) possible scenarios:
1)
IF (and ONLY if) goes to less than 1.4 volts DC:
(page 632).
2) IF REMAINS more than 1.4 volts DC:
The Power Logic Board is bad.
P- 1.0.7 PIN 3 LESS THAN 1.4 VOLTS / ISOLATE FAN AIR OUTPUT i.e. IS THE MACHINE ON?
Figure right, is air felt from the power supply fan vents?
Yes
Fan running!
The machine is on!
No Fan IS NOT running! TWO (2) possible scenarios. Is the screen:
1)
On? OR
2)
Remaining black?
1)
IF (and ONLY if) the screen is on:
The fan is bad!
2)
IF the screen is remaining black:
Unplug the machine and
replace Main Relay K1
*
AND its four pin cable.
*
To LOCATE K1 and its cable refer to
(page 651).
.
P- 1.0.8 FAN RUNNING / ISOLATE INTERMITTENT POWER ON PROBLEM
a) ENSURING the fan continues to run, if (and ONLY if) the screen REMAINS or turns BLACK after forty
(40) seconds see (ABOVE) procedure number
(page 597).
b) ENSURING the fan continues to run, allow adequate time to see if the machine turns itself off. If (and
ONLY if) it DOES turn off proceed to
page 647
, procedure number
c) If the screen remains on address any alarm banners that appear.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...