2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 53
F- 6.1.2 PDIA AND ADIA STABLE BETWEEN 2.0 AND 7.5
A)
Turn the machine OFF!
B) Preferably from another machine, that is not having Flow Errors, swap
in a DiaSafe
®
filter!
C) Turn the m
achine on and return to Dialysis Program (“Select Program” → ‘Dialysis’ →
‘
Enter’)!
D)
OPEN the shunt door till instructed OTHERWISE!
E) Allow five (5) minutes OR until NO AIR bubbles are seen flowing through the DiaSafe housing tubing.
F) From the Home screen, press the [Dialysate Flow] window.
G) Set [Dialysate Flow] to 800 ml/min and press ‘Enter’.
H) If a TMP alarm is NOT PRESENT skip to part J. If a TMP alarm is present:
1)
Press and release the
‘Reset’ key then immediately press and hold it for three (3) seconds;
2)
Allow thirty (30) seconds;
3)
If
a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to part I.
I) Allow thirty (30) seconds BEFORE continuing to part J.
J) Call debug screen 0. WITHOUT LOOKING AWAY watch
Flow Error
for four (4) minutes OR until it
EVER = 1, indicating a Flow Error. TWO (2) possible scenarios 1) or 2) below:
1)
IF (and ONLY if) Flow Error ALWAYS = 0:
Possibly the
Diasafe
®
filter fixed the problem OR the
flow error is intermittent! Proceed to
page 36
, procedure number
2)
IF Flow Error EVER = 1:
Call debug screen 4.
PDIA
should REMAIN between 1.5 and 8.0 AND
STABLE i.e. DOES NOT change more than 0.5 per minute. Watch it for two (2) minutes. TWO (2)
possible scenarios 3) or 4) below:
3)
IF (and ONLY if) PDIA is IS NOT remaining between 1.5 and 8.0 and / or is IS NOT
stable:
(page 53).
4) IF PDIA is STABLE between 1.5 and 8.0:
Close the shunt door THEN proceed to
page 54
, procedure number
F- 6.1.3 PDIA NOT REMAINING BETWEEN 1.5 and 8.0 OR UNSTABLE
a)
CLOSE the shunt door!
b) A procedure, in a different Section, is performed next.
Before continuing to part c
NOTE
this page and
procedure number (F- 6.1.3) as you may be prompted to return here.
c) BEFORE continuing to part d, proceed to
page 527
, to perform
d) If a leaking balancing chamber valve was NOT located in part c, proceed to
page 65
, procedure
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...