2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 65
F- 7.0.4 FLOW ERROR = 1 / ISOLATE ‘FRESH SIDE’ FLOW PATHS
a)
Plug the acid connector into its rinse port to call
the “Select Program” screen but DO NOT press
any keys!
b) Balancing Chamber valve connectors will be switched to test ‘fresh side’ flow paths. Flow is measured
into a 1000 ml
GRADUATED
cylinder for one (1) minute. Good flow = more than 800 ml every minute!
c) ENSURING “Select Program” REMAINS up, perform PARTS 1 AND 2 EVEN IF PART 1 yields bad
flow!
PART 1:
a)
Remove the RED DIALYZER connector from the shunt door and place it into a bucket!
b)
P
lug valve #35’s connector into valve #30’s distribution board position, “V30”.
Leave
valve #30 unplugged!
c) Hold the dialyzer connector up and watch flow (if any) for one (1) minute. If good initially
but significantly slows indicates an intermittent bad valve #31 or #35. If no flow, ENSURE
valve #35’s connector is placed properly in position “V30”. If okay see part d!
d) Measure flow, from the dialyzer connector, into the cylinder, for one (1) minute.
e)
Return valve #35’s connector to its distribution board position!
f) Record Part 1’s measurement
THEN perform PART 2!
PART 2:
a)
P
lug valve #37’s connector into valve #30’s distribution board position, “V30”.
b) Hold the dialyzer connector up and watch flow (if any) for one (1) minute. If good initially
but significantly slows indicates a bad valve #33 or #37. If no flow, ENSURE valve
#37’s connector is placed properly in position “V30”. If okay see part d.
c) Measure flow, from the dialyzer connector, into the cylinder, for one (1) minute!
d) Return the connector to the shunt door.
e) Leaving valve #37 in valve #30’s position for now, analyze BOTH Part 1 AND Part 2’s
Table 2 – Parts 1 AND 2
PART 1’s
measurement
PART 2’s
measurement
Your Response
Less than 800 ml
Less than 800 ml
(page 71)
Less than 800 ml
More than 800 ml
(page 67)
More than 800 ml
Less than 800 ml
(page 69)
More than 800 ml
More than 800 ml
(page 75)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...