2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 111
F- 13.0.0 ‘BOB’ IS MOVING WHEN IT’S NOT SUPPOSED TO
The
Temperature
AND / OR the
Conductivity
window is REMAINING RED BUT the external flow
indicator’s bob is moving!
a) Per the Figure BELOW, trace the wires from Valve #24’s AND #26’s distribution board connectors, to
Valve #24 and #26 in the hydraulics to
ENSURE
they are wired to the CORRECT valve!
b) Figure left, open the plastic female cover from Valve
#26’s distribution board connector.
c) Valve #26 wires MUST be soldered between
pins two
(second from the top) AND five
(bottom). If between the
top and bottom pins Valve #26 is probably connected to
distribution board position “V24” and vice versa!
d) Return Valve #26’s connector to distribution board
position “V26”.
e) ENSURING [Dialysate Flow] is ON (Flow on/off lamp NOT blinking) AND is set to 800 ml/min is he
external flow indicator’s ‘bob’ still rising?
Yes
‘Bob’ moving! TWO (2) possible bad components:
1)
Actuator-Test Board OR;
2)
Valve
#24.
No
‘Bob’ NOT moving! If a Flow Error is still occurring, return to (ABOVE) procedure number
page 50
). If the Flow Error never occurs do not continue.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...