2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 646
P- 2.0.8 ISOLATE ACTUATOR-TEST BOARD ‘SET’ SIGNAL
a)
To avoid damage, turn the machine OFF!
b) Swap in a known good Actuator-Test Board
*
.
*
To LOCATE the Actuator-Test board refer to
c) Turn the machine on and allow forty (40) seconds. Does System
Initialization reach 100%?
Yes
System Initialization reaches 100%! The previous Actuator-Test Board is bad.
No
System Initialization does DOES NOT reach 100%! The previous Actuator-Test Board is
good. TWO (2) possible bad components:
1)
Functional Board
1
;
2)
Motherboard
1
To avoid “Cond Offset Failure”, place the machine into T and C mode! Refer to
OPERATING MODES
(page
Error! Bookmark not defined.
)
LEFT BLANK INTENIONALLY