DAIR- 5.0.0 ISOLATE THE DEAERATION PUMP HEAD
a)
IMPORTANT
!
Remove the clamp from Valve #39.
b) Press ‘CONFIRM’ twice to return to the “Calibrate Hydraulics”
menu.
c) Select Deaeration Pressure but
DO NOT
press any keys.
The
[PUMP RATE]
box MUST = 0 before continuing to part d.
d) Referring
(page 500) AND the Figure right,
swap in a KNOWN good deaeration pump head,
e) Can
[PUMP RATE]
now be adjusted to achieve
target pressures?
Yes
Target pressures remain!
A)
Press CONFIRM to save the calibration;
B)
Turn the machine
off;
C)
Turn the machine on and return to Dialysis Program (“Select Program” → ‘Dialysis’
→ CONFIRM);
D)
The Deaeration problem is solved but return to the procedure that
prompted you here.
No Target pressures CANNOT be achieved! See procedure number
(page 502).
NOTE: The previous pump head may be good!
DAIR- 6.0.0 ISOLATE THE DEAERATION MOTOR
a) Press ‘CONFIRM’ twice to return to the “Calibrate Hydraulics” menu.
b) Select “Deaeration Pressure” but
DO
NOT
‘CONFIRM! The
[PUMP RATE]
data box MUST = 0
before continuing to part c
c) Swap in a KNOWN good deaeration motor.
d) ‘Sharply’ press ‘CONFIRM’. Can
[PUMP RATE]
be adjusted to achieve
target pressures?
Yes
Target pressures achieved! Target pressures achieved!
A)
Press ‘CONFIRM’ to save the
calibration!
B)
Turn the machine off then on and return to Dialysis Program (“Select
Program” → ‘Dialysis’ → ‘CONFIRM’).
C)
The Deaeration problem is solved but return to
the procedure that prompted you here!
No Target pressures CANNOT be achieved! See parts THROUGH d below:
a) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
b) BEFORE continuing to part c, a procedure, in a different Section, is performed.
IMPORTANT!
NOTE
this page and procedure number (DAIR- 6.0.0) because you may
prompted to return to here.
c) Before continuing to part d, proceed to
page 139
d) If a Hydrochamber leak WAS NOT located in part c, THREE (3) possible bad
components:
1)
Bad deaeration motor OR;
2)
Bad deaeration restrictor* (*to LOCATE
refer to
3)
Bad Hydrochamber.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...