2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 443
PDIA OR ADIA is
IS NOT
between 3.0 and 6.0 continued:
b) Continuous flow, more than 0.2 ml per minute, from the connector?
Yes Continuous flow!
A)
Return the connector to the shunt and close the door;
B)
Turn the machine OFF;
C)
Proceed to
page 527
No continuous flow! See parts a THROUGH c below:
a)
Return the connector to the shunt AND close door!
b) Call debug screen 0 to ENSURE the TOP balancing chamber valves
are REMAINING BLUE!
c) Watch Valve #41’s ‘dot’ (Figure right) for one (1)
minute. It should stay white! TWO (2) possible
scenarios:
1)
IF (and ONLY if) Valve #41’s ‘dot’ REMAINS white!
Proceed to
page 462
, procedure number
2) IF Valve #41’s ‘dot’ cycles between white and blue:
A)
Turn
the machine OFF!
B)
Proceed to
page 527
TMP- 1.0.2 PDIA AND ADIA STABLE BETWEEN 3.0 AND 6.0
a)
IMPORTANT!
At the bottom of the screen, select the ‘Dialysate’ tab.
b) Press ‘CONFIRM’.
c) Figure right, if (and ONLY if) the screen’s ‘status’ bar
is NOT BLANK see procedure number
(page 444). If BLANK repeat parts a and b!
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...