2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 431
FIL- 4.0.0 FILACT = 0 / FINAL ‘FILLING PROGRAM’ CHECKS
Call the Home screen. Figure right, if the
TMP
window is RED a TMP alarm
is present. TWO (2) possible scenarios:
1)
IF (and ONLY if) the TMP window is WHITE:
2)
IF the TMP window is RED:
Perform parts a THROUGH d below:
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b)
IMPORTANT!
Allow thirty (30) seconds!
c) If a TMP alarm reoccurs perform parts a and b up to twice more BEFORE continuing to part d.
d) Allow thirty (30) seconds THEN see procedure number
(page 431).
FIL- 4.0.2 ISOLATE SPENT SIDE BALANCING CHAMBER LEAK
TMP is STABLE if the TMP window REMAINS white AND does NOT change more than 60 mmHg in three
(3) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is STABLE:
ASSUMING all procedures were performed correctly (i.e. no
air seen at Chamber #69) AND if (and ONLY if)
FILACT
continues to = 1 intermittently, see
(page 430).
2) IF TMP is UNSTABLE:
See parts a THROUGH c below:
a) A procedure, in different Section, is performed next.
NOTE
this page and procedure number
(FIL- 4.0.2) as you may prompted to return here. Perform parts a and b below
b) Before continuing to part b, proceed to
page 527
, to perform
c) If balancing chamber valve leak was NOT located in part a, proceed to
page
SECTION 9- TMP (PRESSURE) PROBLEMS
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...