2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 100
Yes
More than 800 ml!
A)
Reconnect Valve #24’s tubing;
B)
NOTE
ONLY
VALVE #24
will be checked and proceed to
No TWO (2) possibilities:
1)
Bad (restricted) DiaSafe
®
filter OR
2)
Tubing restriction
between the balancing chambers and Valve #24.
F- 9.0.43 GOOD FLOW FROM VALVE #24
A restriction is indicated in the ‘from (red) Dialyzer line’.
a) ENSURE no tubing restrictions at the DiaSafe
®
filter.
b) Figure BELOW, open Dialysate Filter #73’s housing and ENSURE the filter is clean!
c) Figure above touch Valve #25’s black solenoid. Is it warm?
Yes
Solenoid warm! TWO (2) possible bad components: 1) Bad Actuator-Test Board
1
; 2) Bad
Valve #25.
1
To LOCATE the board refer to
(page 9).
No Solenoid is not warm!
NOTE
ONLY
VALVE #25
will be checked and proceed to
page 192