2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 276
c) Per the Figure below, measure at the back side of the SENSOR BOARD’S X2
connector at
pin 13
(bottom row, 7 pins from the rear of machine).
d) Figure right, does the meter’s numeric AND units display
read more than 12,000
Ω
(12.0
KΩ
)
?
Yes
More than 12,000 Ω (12.0
KΩ
)! Assuming all procedures were
performed correctly and the “no voltage to the heater” problem
continues, see parts a through c below:
a)
Swap in a
known good
Power Logic Board!
b)
Place the machine into Dialysis Program or Heat Disinfect!
c)
Measure for DC voltage again at the Sensor Board’s motherboard
connector pin C1 (VERY front pin). A good signal is 3.6 volts DC or
less. If a bad signal is measured the motherboard may be bad.
No
Less than 12,000 Ω! See parts A THROUGH C below:
A) ENSURE the machine was OFF prior to the measurement!
B) Referring to the Figure below, unplug the Sensor Cable.
C) Repeat the measurement at the SENSOR BOARD’S X2 connector at
pin 13
. More than 12,000
Ω (12
K
Ω
) now?
Yes
More than 12,000 Ω. TWO (2) possible bad components;
1)
Bad Sensor Board cable OR;
2)
Bad distribution board.
No The Sensor Board is bad
Figure 46 – Sensor Board / NTC #2 Check
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...