2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 486
PHT- 4.4.0 REPEAT PRESSURE TESTS
Call debug screen 1. While watching the screen banners
1
,
ATMP
1
AND
FILACT
2
allow the tests to finish.
Do BOTH (Negative and Positive) pass?
Yes
BOTH pass!
IMPORTANT!
(page 486).
No Pressure Test Failed! Return to (ABOVE) procedure number
(page 481) to analyze the
failure(s).
1
If
ATMP
changes more than 20 during Neg Flow On’s HOLD period indicates a leak
2
After “Remove Air”, if
FILACT
goes to 1 the indicates an air leak.
PHT 4.5.0 BOTH TESTS PASS / PERFORM UF INTEGRITY TEST
The Important UF INTEGRITY TEST, performed here, isolates the UF Pump for leaks:
a)
Figure right
, remove the TOP
tubing from between the UF Pump and
Check Valve #63.
b) Attach a tubing segment to the vacant
Pump nozzle that is long enough to route
it away from the hydraulics.
c) RESET all alarms!
d) From the
[Test & Options]
screen
repeat the Pressure Tests.
e) Do BOTH Pressure Tests pass?
Yes
BOTH Tests pass! See procedure number
No a) Rebuild the UF Pump per the
Preventative Maintenance
Procedures booklet
ENSURING the springs and
seals are properly installed!
b) Return to Dialysis Program.
c) Repeat the Pressure Tests.
They MUST pass with AND
without the UF Pump OUTPUT
tubing attached!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...