2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 177
CLEAN- 4.3.2 VERR = 0 / VERIFY TOTAL FLOW
Watching for fifteen (15) seconds, does the external flow indicator’s ‘bob’ EVER rise at least ¼ way up in
the sight tube?
Yes
‘Bob’ moving! Proceed to
page 180
, procedure number
No
‘Bob’ NOT moving! See parts a THROUGH d below:
a) ENSURE no tubing restrictions at the Flow Pump and DiaSafe
®
filter!
b) ENSURE Flow Pump Pressure (gauge at Flow Pump) is still PEAKING to between 34
and 38 psi.
c) ENSURE Loading Pressure (gauge in Rinse port) PEAK pressure remains between 22
and 27 psi
d) A restriction is indicated! See procedure number
CLEAN- 4.4.0 ‘BOB’ NOT MOVING / ISOLATE VALVE #29
a) Screw a 60 ml syringe, COMPLETELY filled with water,
onto the Fluid Sample Port (Figure right).
b) Start a SIX (6) second timer, in your head, as you
begin to push
AS HARD AS YOU CAN
on the
syringe plunger! You should feel very little resistance!
c) Can you push ALL of the water through the Sample
Port within SIX (6) seconds?
Yes
Very little resistance encountered! Valve #29 is
open! See procedure number
No
Significant resistance encountered!
NOTE
ONLY
VALVE #29
will be checked and proceed to
page 192
CLEAN- 4.4.1 VALVE #29 OPEN / ISOLATE VALVE #24
a) A 1000 ml (or larger) graduated cylinder is required below.
b) ENSURE the water is ON!
c)
Open the shunt door and LEAVE it OPEN until instructed!
d) Remove the BLUE dialyzer connector but
DO NOT CLOSE THE DOOR!
e) Measure from the dialyzer connector for thirty (30) seconds! More than three-hundred (300) ml
collected?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...