SECTION 27 - “ACT BLOOD PUMP FAILED”
A) Turn the machine off and back on. If (and ONLY if) “Act Blood Pump Failed” alarm reoccurs proceed to
part B.
B)
IMPORTANT!
(page 6)!
(page 661).
ABPF- 1.0.0 TROUBLESHOOT ACT BLOOD PUMP FAILED
a) Turn the machine OFF and, to prevent pulling cables loose, GENTLY open the card cage!
b)
IMPORTANT! Set your CALIBRATED volt meter to Volts DC (V
DC
).
c)
Connect the meter’s ground (black) lead to chassis ground (see
, page
d) Spread the card cage side panels open then gently drop the front panel down to access the
motherboard’s nine (9) pin TEST Connector (Figure below).
e) TIGHTLY HOLD the red meter lead at the TEST Connector’s
+24V-C
pin (pin 9).
f) Looking for approximately 24.0 volts at least once, WITHOUT LOOKING AWAY
from the meter, for thirty
(30) seconds,
turn the machine on (fan running)!
More than 23.0 volts DC
(V
DC
) seen
at ANY
TIME (Yes or No)?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...