2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 234
T- 1.0.22 TEMPERATURE LESS THAN
35.1° C
a)
Set your CALIBRATED volt meter to AC volts
(~ V, V
AC
)!
b) Figure right, unplug the connector from the
1
st
distribution
board position from the left, “CON-NTC”. This is NTC #2.
c) Figure right, measure at the distribution board’s Heater
Connector, between the BROWN and BLUE wires.
d) TWO (2) possible scenarios:
1)
IF (and ONLY if) less than 100.0 volts AC:
Leaving
NTC #2 unplugged till instructed, proceed to
page
, procedure number
2) IF MORE THAN 100.0 volts AC:
Leaving NTC #2
unplugged, see parts a AND b below:
a) ENSURE the external flow indicator’s ‘bob’ is moving up and down.
b) WITHOUT LOOKING AWAY, watch
the meter for UP TO SEVEN (7) minutes
OR
until if it
EVER reaches more than 35.5° C even if only once?
Yes
35.5° C OR more!. See procedure number
No Less than 35.5° C! Either incoming water is EXTREMELY cold OR you made an error
and heater voltage IS NOT more than 100 volts AC OR the heater is intermittent bad.
T- 1.0.23 MEASURED TEMPERATURE, WITH NTC #2 UNPLUGGED, EXCEEDS 35.5° C
a)
IMPORTANT!
Return NTC #2’s connector to
1
s
t
distribution board position from the left, “CON-NTC”.
b) If a
6.04 KΩ plug was previously placed in NTC #3’s distribution board position (“MON-NTC”) remove it
and return NTC #3’s connector to the
2
nd
position from the left!
c) Call debug screen 4 to see
TEMP
(lower left).
TWO (2) possible scenarios:
1)
IF (and ONLY if) 4.0 or more:
NTC #3 is okay. See procedure number
(page 234).
2) IF LESS THAN 4.0:
NTC #3* is plugged in incorrectly OR is bad. *To LOCATE NTC #3 refer to
(page 218).
T- 1.0.24 TEMP IS 4.0 OR MORE
a) If a 274
Ω plug was previously placed at the Cond Cell’s distribution board position, “X7, COND”, remove
it and return the Cond Cell’s connector.
b) Proceed to
page 260
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...