2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 221
T- 1.0.7 TEMPERATURE PROBLEM
A) Remove a ‘dummy chamber’ from the Level Detector module.
B) Connect to known good acid
*
and LIQUID bicarbonate
*
!
*
Tested per clinic procedures!
C)
Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM).
D) From the Home screen, select the [Dialysate Flow] window.
E) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
F) Call debug screen 0. If
Flow Error
EVER = 1, even just once, indicates a Flow Error. WITHOUT
LOOKING AWAY, watch it for two (2) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error ALWAYS = 0:
See part G.
2)
IF (and ONLY if), Flow Error EVER = 1:
Proceed to
page 20
G) Call debug screen 6. If
BC Switch
(middle column) is EVER = 897 or more, even
just once, indicates a Flow Error. Watch
it for two (2) minutes or until if it EVER =
897 or more. TWO (2) possible scenarios:
1)
IF (and ONLY if), BC Switch EVER = 897 or more:
Proceed to
page 20
2)
IF BC Switch is NEVER, EVER = 897 or more:
Call the Home screen to see [
Temperature
].
TWO (2) possible scenarios:
Scenario #1:
IF (and ONLY if) remains less than 35.1° C:
Proceed to
page
, procedure number
Scenario #2:
IF EVER more than 35.1° C:
(page 222).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...