2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 18
h) The "CONFIRM Concentrate” screen can be called at any time by selecting the screen’s lower
[Dialysate] tab or the Home screen’s [Conductivity] window.
Auto Flow
Selecting Dialysate Flow rates more than 800 ml/min enables
auto flow
(1.5x or 2x). At these settings
Dialysate Flow rate will automatically adjust itself, in 100 ml/min increments, depending on blood pump
rate. For example, at 2X, when the blood pump rate is set to 250 ml/min, Dialysate Flow will set itself to
500 ml/min (2 x 250) and the Home screen’s [Dialysate Flow] window will display preceding “a” (i.e.
a500”).
Do NOT select 1.5x or 2x while troubleshooting!
End of ‘How to Enter Dialysis Program’
How to Enter a Cleaning / Disinfection Program:
a) Return both dialyzer connectors to the shunt and close the door.
b) Place both concentrate connectors into their rinse ports. This calls the “Select Program” screen
(1)
.
c) Select the desired Cleaning / Disinfection Program button then follow screen prompts that may appear
depending on what Cleaning Program is running.
End of ‘How to Enter a Cleaning / Disinfection Program’
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...